(b) Classification. Class I (general controls). The device is exempt from the premarket notification procedures in Subpart E of Part 807. The device also is exempt from the good manufacturing practice regulation in Part 820 with the exception of § 820.180, with respect to general requirements concerning records, and § 820.198, with respect to complaint files.

§ 880.6230 Tongue depressor.

(a) Identification. A tongue depressor is a device intended to displace the tongue to facilitate examination of the surrounding organs and tissues.

(b) Classification. Class I (general controls). The device is exempt from the premarket notification procedures in Subpart E of Part 807. If the device is not labeled or otherwise represented as sterile, it also is exempt from the good manufacturing practice regulation in Part 820, with the exception of § 820.180, with respect to general requirements concerning records, and § 820.198, with respect to complaint files.

§ 880.6250 Patient examination glove.

(a) Identification. A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

(b) Classification. Class I (general controls).

[45 FR 69682-69737, Oct. 21, 1980, as amended at 53 FR 1604, Jan. 13, 1989]

§ 880.6265 Examination gown.

(a) Identification. An examination gown is a device intended for medical purposes that is made of cloth, paper, or other material that is draped over or worn by a patient as a body covering during a medical examination.

(b) Classification. Class I (general controls). The device is exempt from the premarket notification procedures in Subpart E of Part 807. If the device is not labeled or otherwise represented as sterile, it also is exempt from the good manufacturing practice regulation in Part 820, with the exception of § 820.180, with respect to general requirements concerning records, and

§ 820.198, with respect to complaint files.

§ 880.6280 Medical insole.

(a) Identification. A medical insole is a device intended for medical purposes that is placed inside a shoe to relieve the symptoms of athlete's foot infection by absorbing moisture.

(b) Classification. Class I. The device is exempt from the premarket notification procedures in Subpart E of Part 807 of this chapter.

[45 FR 69682-69737, Oct. 21, 1980, as amended at 54 FR 25050, June 12, 1989]

§ 880.6320 AC-powered medical examination light.

(a) Identification. An AC-powered medical examination light is an ACpowered device intended for medical purposes that is used to illuminate body surfaces and cavities during a medical examination.

(b) Classification. Class II (performance standards).

§ 880.6350 Battery-powered medical examination light.

(a) Identification. A battery-powered medical examination light is a battery-powered device intended for medical purposes that is used to illuminate body surfaces and cavities during a medical examination.

(b) Classification. Class I (general controls). The device is exempt from premarket notification procedures in Subpart E of Part 807. The device also is exempt from the good manufacturing practice regulation in Part 820, with the exception of § 820.180, with respect to general requirements concerning records, and § 820.198, with respect to complaint files.

§ 880.6375 Patient lubricant.

(a) Identification. A patient lubricant is a device intended for medical purposes that is used to lubricate a body orifice to facilitate entry of a diagnostic or therapeutic device.

(b) Classification. Class I (general controls).

§ 880.6430 Liquid medication dispenser.

(a) Identification. A Liquid medication dispenser is a device intended for

medical purposes that is used to issue a measured amount of liquid medication.

(b) Classification. Class I (general controls). The device is exempt from the premarket notification procedures in Subpart E of Part 807. The device also is exempt from the good manufacturing practice regulation in Part 820, with the exception of § 820.180, with respect to general requirements concerning records, and § 820.198, with respect to complaint files.

§ 880.6450 Skin pressure protectors.

(a) Identification. A skin pressure protector is a device intended for medical purposes that is used to reduce pressure on the skin over a bony prominence to reduce the likelihood of the patient's developing decubitus ulcers (bedsores).

(b) Classification. Class I (general controls). The device is exempt from the premarket notification procedures in Subpart E of Part 807. The device also is exempt from the good manufacturing practice regulation in Part 820, with the exception of § 820.180, with respect to general requirements concerning records, and § 820.198, with respect to complaint files.

§ 880.6500 Medical ultraviolet air purifier. (a) Identification. A medical ultraviolet air purifier is a device intended for medical purposes that is used to destroy bacteria in the air by exposure to ultraviolet radiation.

(b) Classification. Class II (performance standards).

§ 880.6710 Medical ultraviolet water purifier.

(a) Identification. A medical ultraviolet water purifier is a device intended for medical purposes that is used to destroy bacteria in water by exposure to ultraviolet radiation.

(b) Classification. Class II (performance standards).

§ 880.6730 Body waste receptacle.

(a) Identification. A body waste receptacle is a device intended for medical purposes tht is not attached to the body and that is used to collect the body wastes of a bed patient.

[blocks in formation]

§ 880.6760 Protective restraint.

(a) Identification. A protective restraint is a device, usually a wristlet, anklet, or other type of strap, that is intended for medical purposes and that limits a patient's movements to the extent necessary for treatment, examination, or protection of the patient.

(b) Classification. Class I (general controls). The device is exempt from the premarket notification procedures in Subpart E of Part 807. The device is also exempt from the good manufacturing practice regulation in Part 820, with the exception of § 820.180, with respect to general requirements concerning records, and § 820.198, with respect to complaint files.

§ 880.6775 Powered patient transfer device.

(a) Identification. A powered patient transfer device is a device consisting of a wheeled stretcher and a powered mechanism that has a broad, flexible band stretched over long rollers that can advance itself under a patient and transfer the patient with minimal dis

turbance in a horizontal position to the stretcher.

(b) Classification. Class II (performance standards).

§ 880.6785 Manual patient transfer device.

(a) Identification. A manual patient transfer device is a device consisting of a wheeled stretcher and a mechanism on which a patient can be placed so that the patient can be transferred with minimal disturbance in a horizontal position to the stretcher.

(b) Classification. Class I (general controls). The device is exempt from premarket notification procedures in Subpart E of Part 807. The divice also is exempt from the good manufacturing practice regulation in Part 820, with the exception of § 820.180, with respect to general requirements concerning records, and § 820.198, with respect to complaint files.

§ 880.6800 Washers for body waste receptacles.

(a) Identification. A washer for body waste receptacles is a device intended for medical purposes that is used to clean and sanitize a body waste receptacle, such as a bedpan. The device consists of a wall-mounted plumbing fixture with a door through which a body waste receptacle is inserted. When the door is closed the body waste receptacle is cleaned by hot water, steam, or germicide.

(b) Classification. Class I (general controls). The device is exempt from the premarket notification procedures in Subpart E of Part 807. The device also is exempt from the good manufacturing practice regulation in Part 820 with the exception of § 820.180, with respect to general requirements concerning records, and § 820.198, with respect to complaint files.

§ 880.6820 Medical disposable scissors.

(a) Identification. Medical disposable scissors are disposable type general cutting devices intended for medical purposes. This generic type of device does not include surgical scissors.

(b) Classification. Class I (general controls). The device is exempt from premarket notification procedures in Subpart E of Part 807.

[blocks in formation]

platform mounted on a wheeled frame that is designed to transport patients in a horizontal position. The device may have side rails, supports for fluid infusion equipment, and patient securement straps. The frame may be fixed or collapsible for use in an ambulance.

(b) Classification. Class II (performance standards).

§ 880.6920 Syringe needle introducer.

(a) Identification. A syringe needle introducer is a device that uses a spring-loaded mechanism to drive a hypodermic needle into a patient to a predetermined depth below the skin surface.

(b) Classification. Class II (performance standards).

§ 880.6960 Irrigating syringe.

(a) Identification. An irrigating syringe is a device intended for medical purposes that consists of a bulb or a piston syringe with an integral or a detachable tube. The device is used to irrigate, withdraw fluid from, or instill fluid into, a body cavity or wound.

(b) Classification. Class I (general controls). The device is exempt from the premarket notification procedures in Subpart E of Part 807. If the device is not labeled or otherwise represented as sterile, it also is exempt from the good manufacturing practice regulation in Part 820, with the exception of § 820.180, with respect to general requirements concerning records, and § 820.198, with respect to complaint files.

[blocks in formation]

§ 880.6980 Vein stabilizer.

(a) Identification. A vein stabilizer is a device consisting of a flat piece of plastic with two noninvasive prongs. The device is placed on the skin so that the prongs are on either side of a vein and hold it stable while a hypodermic needle is inserted into the vein.

(b) Classification. Class I (general controls). The device is exempt from the premarket notification procedures in Subpart E of Part 807. If the device is not labeled or otherwise represented as sterile, it is also exempt from the good manufacturing practice regulation in Part 820, with the exception of § 820.180, general requirements concerning records, and § 820.198, with respect to complaint files.

[blocks in formation]

Esthesiometer.

882.1525 Tuning fork.

882.1540 Galvanic skin response measurement device.

882.1550 Nerve conduction velocity measurement device.

882.1560 Skin potential measurement device.

[blocks in formation]
[blocks in formation]

882.4150 Scalp clip.

882.4175

Aneurysm clip applier.

882.4190 Clip forming/cutting instrument. 882.4200 Clip removal instrument.

[blocks in formation]

882.5800 Cranial electrotheraphy stimulator.

882.5810 External functional neuromuscular stimulator.

882.5820 Implanted cerebellar stimulator. 882.5830 Implanted diaphragmatic/phrenic nerve stimulator.

882.5840 Implanted intracerebral/subcortical stimulator for pain relief. 882.5850 Implanted spinal cord stimulator for bladder evacuation.

882.5860 Implanted neuromuscular stimulator.

882.5870 Implanted peripheral nerve stimulator for pain relief.

882.5880 Implanted spinal cord stimulator for pain relief.

882.5890 Transcutaneous electrical nerve stimulator for pain relief.

882.5900 Preformed craniosynostosis strip. 882.5910 Dura substitute.

882.5940 Electroconvulsive therapy device. 882.5950 Artificial embolization device. 882.5960 Skull tongs for traction.

AUTHORITY: Secs. 501, 510, 513, 515, 520, 701 of the Federal Food, Drug, and Cosmetic Act (21 U.S.C. 351, 360, 360c, 360e, 360j, 371).

SOURCE: 44 FR 51730-51778, Sept. 4, 1979, unless otherwise noted.